HI-3520 Manual

Page 7

HI-3520 Display

Figure 3

The beep and the flash rate are directly proportional to

the RF field strength. Note that once the alarm starts, it

will latch at the minimum flash rate, continuing to warn

of an over-exposure even if the RF field that caused the

alarm is no longer present. In order to reset the alarm

function, the unit must be turned OFF and back ON using

the ON-OFF switch. For information concerning disabling

the latch function, refer to section 3.11. For information

concerning "Instantaneous" vs "Average" measurement

modes, refer to section 3.8.

3.5 Set-up Mode

The "Set-up" mode is used to set the alarm warning level

and to enable/disable the audible alarm. To enter this

mode, set the "ON-OFF" switch to ON. Then depress the

MODE switch after the initial three second period

referenced in section 3.4. The LCD display will indicate

"S.U.". At this point, the alarm warning level can be

changed (section 3.10) or the audible alarm indicator can

be enabled or disabled (section 3.7).

3.6 Setting Alarm Warning Level