HI-3520 Manual

Page 9

or 100 measurements), the running average of all

measurements taken will be displayed. After six minutes,

each "new" measurement will replace the measurement

taken 363.3 seconds prior to the "new" measurement.

Therefore, the average of the most recent 100

measurements will always be displayed.

The average of the latest 100 measurements is also

compares by the unit to the alarm warning level set by

the operator (section 3.6) to determine the % ALARM

LEVEL ratio. (During the first six minutes of operation,

the running average of all measurements taken will be

compared by the unit to the alarm warning level set by

the operator to determine the % ALARM LEVEL ratio.) If

the % ALARM LEVEL ratio reaches 100%, the unit's

alarm circuitry will be activated, the red LED and the

ALARM annunciator on the LCD will flash every 3.6

seconds, and the audible alarm will sound every 3.6

seconds.

3.10 "Instantaneous" Measurement Mode

When the INST annunciator on the left side of the LCD

display area is visible, the measured power density is

sampled and displayed every 1.8 seconds. (Note that the

Model HI-3520 saves data for six minutes regardless of

the display selected. Switching between "Average" and

"Instantaneous" does affect the data being stored.)

3.11 Latched Alarm

A feature of the unit's alarm system is that the beeps and

flashes are progressive. Once the alarm level has been

reached, the unit will indicate a single beep and flash

each 1.8 second with the "INST" annunciator displayed

or 3.6 seconds with the "AVG" annunciator displayed.

Should the unit detect a field that exceeds the preset

alarm level, the unit will start to beep and flash. The