beautypg.com

Archived 4/2/10 – ETS-Lindgren HI-3520 Microwave Monitor (Archived) User Manual

Page 14

background image

Page 10

HI-3520 Manual

beep and the flash rate are directly proportional to the RF

field strength. Note that once the alarms start, they latch

at the minimum flash rate, continuing to warn of an over

exposure situation even it the RF field that initially caused

the alarm is no longer present. In order to reset the alarm

function, it is necessary to turn the unit OFF and back ON

using the ON-OFF switch. If it is desired to disable the

latch function, follow the turn-on procedure described in

section 3.4 except depress and hold the MODE switch

before placing the ON-OFF switch to the ON position.

3.12 Built-In Test Functions

The unit generates a beep and flash at turn on to verify

that the alarm indicators are functioning. At this time, all

segments of the LCD display are enabled to indicate that

the display is operating properly. Ten seconds after turn

on. the unit will generate a second beep and flash which

indicated that it is operational. A low battery condition

is indicated if the unit beeps and flashes two times after

the ten second period. To verify, check the LCD display

for a "BAT" indication. In addition, the battery condition

is tested automatically every minute. Should the battery

measure low, the "BAT" indicator will flash and the unit

will beep and flash once a minute. Refer to Section 5.0

for battery replacement procedures.

The unit also contains continuous fault detection circuitry

to alert the user to a RF detector failure. In that event,

the beeper and light will start to beep and flash at a

constant tone and light, and the LCD display will indicate

"FAIL". The unit is no longer usable in this condition and

should be turned off and serviced.

Archived 4/2/10