power density level on the test data sheet in

mW/cm

2.

3.0

Repeat step 2.0 at 2.45, 3.8, 8.0, 12.0 and

18.0 GHz.

4.0

From the data in step 2.0 and 3.0, calculate the

average power density measured over the band

by taking the square root of the product of the

highest and lowest readings recorded. Record

the resultant on the data sheet for P

cal

.

5.0

Divide 0.2 mW/cm

2

by the power density level

recorded on the data sheet for P

cal

. This

determines the correction multiplier, M.

6.0

Alignment can now be performed at any one of

the calibration frequencies specified in steps 2.0

and 3.0. Select a convenient calibration

frequency.

7.0

At the selected calibration frequency, multiply by

M the indicated power density from the data

recorded in step 2.0 and 3.0 above. Record this

as P

REF

.

8.0

Remove the battery compartment cover and

located the calibration potentiometer behind the

hole adjacent to the battery cover screw hole.

NOTE:

It is necessary to remove the battery cover door
to adjust the calibration potentiometer but the
cover must be replaced for the RF test.

9.0

Apply the P

REF

power at the calibration

frequency. Adjust this potentiometer such that

Archived 4/2/10