6.0 CALIBRATION

NOTE

The following is applicable only to units with
serial numbers higher than 220104. Please
contact the Holaday Industries Inc. for calibration
instructions for units with serial numbers of
220104 and lower.

6.1 General

The Model HI-3520 Radiation Badge should be calibrated

at 12 month intervals. To do so requires the use of

highly specialized test equipment and facilities, such as a

radio-frequency anechoic chamber and test cells, wherein

fields of known power density can be accurately

established. It should also be noted that to establish the

required power densities will require CW sources of 10W

or greater together with accurately calibrated test

antennae and power monitoring equipment. If facilities of

this type are not available, the unit should be returned to

General Microwave or to another qualified calibration

facility for this service.

6.2 Calibration Procedure

Set the alarm warning level to 0.2 mW/cm

2

as described

in section 3.6.

Disable the latch function as described in section 3.11.

1.0

Position the unit in an anechoic chamber. The

preferred orientation of the E-field is parallel to

the vertical axis of the unit.

2.0

At 1.4 GHz, raise the power density level at the

unit until an alarm is indicated. Record this

Archived 4/2/10