In the "Set-Up" mode (section 3.5), depress and hold the

MODE switch. After three seconds, the preset alarm

warning level will be displayed. Release the MODE

switch. Depressing and holding the MODE switch again

at this point will result in a rapidly increasing alarm

warning level indication. Continue to depress the MODE

switch until the desired alarm level is approached. Then

release the MODE switch and pulse it at a convenient

rate until the desired alarm warning level is reached.

Once it is reached, set the ON-OFF switch to OFF and

then to ON again. The alarm warning level that has just

been set will remain in memory until changed again as

described above.

3.7 Enabling/Disabling Audible Alarm

In the "Set-Up" mode (section 3.5), continue to depress

and release the MODE switch and observe that the

audible alarm indicator (a small bell in the upper left area

of the LCD display) appears and disappears. If an audible

alarm is desired, allow the audible alarm indicator to

remain visible; if an audible alarm is not required,

continued to depress and release the MODE switch until

the audible alarm indicator disappears. Using the ON-OFF

switch, turn the unit OFF and then ON again.

3.8 Measurement Modes

The Model HI-3520 Measurement Mode can be set to

either "Instantaneous" or "Average" by depressing and

releasing the MODE switch while the unit is in its normal

operating mode. See sections 3.9 and 3.10.

3.9 "Average" Measurement Mode

When the AVG annunciator on the left side of the display

area is visible, a six-minute running average is maintained

by the unit. Power density is sampled every 3.6

seconds. During the first six minutes (i.e.-360 seconds