CALIBRATION
CHAPTER 5
Example: 08.0622 mV
7.
Program the measured voltage into the SURETEST Indicator as demonstrated in the example,
“+08.0622” by pushing and holding the Up (
↑) or Down (↓) Cursor keys.
8.
Press ENTER (
↵). The display momentarily reads:
Saved.
9.
Turn the power to OFF. Remove the calibration fixture.