Calibration cw – CDI Torque SURETEST AND 600TL Torque Calibration System User Manual
Page 51

CALIBRATION
CHAPTER 5
6.
Enter the following code into the display using the Cursor Keys:
5222
( Left (
←), Right (→) to select digit ; Up (↑), Down (↓) = increment/decrement value ; Enter (↵) to accept )
Calibration CW:
7. Select
Calibrate
CW. Press ENTER (
↵).
A warning is displayed.
Select ‘
Yes
’. Press
Enter
(
↵) .
8. Using the Up (
↑) or Down (↓) Cursor keys, enter “
1
” for a
single transducer,
or enter “
4
” for the
4-in-1
transducer.
9.
Press ENTER (
↵). The display reads:
Select transducer by applied torque on it, LED will lit.
5-8