6 display software and hardware version, 7 display test, 8 keyboard and buzzer test – Salter Brecknell 6700U Series (USB) User Manual
Page 33

67XXU USB/Serial Scale Service Manual
33
The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TM. If the
time of scale variation being over USER-OTHER-NLD.RG is more than USER-HOLD-
STB.TM,
STB.ER will be shown. Press the [TARE] key to start averaging again or
press the [HOLD] key to exit.
4.6 Display Software and Hardware Version
Display the current version of software and hardware.
1.
Enter the Setup Mode by pressing the [HOLD] + [ON/OFF] keys until
confg
is displayed.
2.
Use the [UNIT] key to select the MISC-VER menu.
3.
Press [TARE] to display the current version.
3a.
Firmware Version display Format is: Vxx.yy, xx is hardware version, yy is
software version.
4.
Press the [ON/OFF] key to return to last menu item.
4.7 Display Test
1.
Press the [HOLD] + [ON/OFF] keys until
confg is displayed.
1.
Use the [UNIT] key to select TEST-DSP.TS menu.
2.
Press [TARE] to enter the test display mode and all segments will light at first.
2a.
In this mode, every press of the [UNIT] key will light the next segment. Every
press of the [HOLD] key will light the next digit.
2b.
Press the [TARE] key to automatically light all segments and all digits.
3.
Press the [ON/OFF] key to return to last menu item.
4.8 Keyboard and Buzzer Test
1.
Press the [HOLD] + [ON/OFF] keys until
confg is displayed.
2.
Use the [UNIT] key to select TEST-KEY.TS menu.
3.
Press [TARE] to enter test keypad mode and
key will be displayed.
4.
Press any key. The value of this key will be displayed and the buzzer will beep.
5.
Press the [ON/OFF] key to return to last menu item.