Salter Brecknell CS-250 User Manual
Page 14

CS-250 Series Chair Scales Service Manual
- 13 -
F2 COM
P cnt2
Another mode, to send data continuous,
P STAB
Send data when the display stable
F3 bk
Bl on
Set the backlight always on.
Bl off
Set the backlight always off.
Bl au
Set the backlight automatic on.
F4 str
Str on
Multi tare operation turn on
Str of
Multi tare operation turn off
tch
pin
Press password to enter into the technical
parameter
P1 spd
Spd 7.5
To select display AD speed,
Spd 15
Spd 30
Spd 60
P2 cal
Cur ut
kg
Select to current weighing unit “kg”
Cur ut lb Select to current weighing unit “lb”
Cal ut
kg
Select to calibrati
on unit “kg”
Cal ut lb Select to calibration unit “lb”
desc
To select scale decimal points;
Options:
C 0, C 0.0, C 0.00, C0.000
C0.0000
Inc
To select scale division/increment;
Options:
Div 1, Div 2, Div 5, Div 10,
Div 20, Div 50
Cap
Select to set scale capacity
Cal
Scale Calibration; details check the calibration
section
tri
To modify the calibration.
This display will be show XXXXX. For trimming
the load cells, showing primary weight.
You can calculate new rate by this formula:
N2=N1+N1×[(K2-K1)÷K2]
N1: primary rate, N2: new rate, K1: calibrate
weight, K2: display weight
count
To show the scale internal count
reset
Reset the scale
setgra
Set the local gravity value
Note: When Jumper K2 is connected on the PCB, then only can access
“tch” parameters.