Page C-1

NAME/CODE

DESCRIPTION

CODE/VALUE

F9
Display
Divisions

Determines the desired weight increments. Value should be
consistent with legal requirements.

1

2
5

F10
Decimal Pt.

Determines location of the decimal point.

0

√ 0.0

0.00 0.000
0.0000 00

F11
Custom Unit
Multiplier

Actuates the function that allows you to enter a multiplier for the
custom unit. Scrolling down with the ZERO key one level begins the
procedure.

Press ZERO key
to begin sequence

1

F12
Custom Unit
Enable

Used to enable or disable the custom unit whenever the UNITS key is
pressed. This feature must be disabled for commercial applications.
"0" = Disable custom unit "1" = Enable custom unit

0
1

F13
Piece Count
Enable

Used to enable or disable the piece count feature. This feature must
be disabled for commercial applications.
"0" = Disable piece count "1" = Enable piece count

0
1

F14
lb/kg
Conversion
Enable

Used to enable or disable the lb to kg conversion whenever the
UNITS key is pressed.
"0" = Disable conversion "1" = Enable conversion

0
1

F16
Zero
Calibration

Places indicator into the zero calibration routine. Scrolling down with
the ZERO key one level begins the procedure.

Press ZERO key
to begin sequence

F17
Span
Calibration

Places indicator into the span calibration routine. Scrolling down with
the ZERO key one level begins the procedure.

Press ZERO key
to begin sequence

F18
View Calibra-
tion

Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the ZERO key one level begins the procedure.

Press ZERO key
to begin sequence

F19
Key-in Zero

Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the ZERO key one level begins
the procedure.

Press ZERO key
to begin sequence

F20
Key-in Span

Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the ZERO key one level
begins the procedure.

Press ZERO key
to begin sequence

F21
Factory Reset

This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!

Press the ZERO
key twice to
execute.