Salter Brecknell 400ES User Manual

Page 14

background image

Page 4 - 2

NAME/CODE

DESCRIPTION

CODE/VALUE

F9
Display
Divisions

Determines the desired weight increments. Value should be
consistent with legal requirements.

1

2
5

F10
Decimal Pt.

Determines location of the decimal point.

0

0.0

0.00

0.000

0.0000

00

F16
Zero
Calibration

Places indicator into the zero calibration routine. Scrolling down with
the #2 key one level begins the procedure.

Press #2 key to
begin sequence

F17
Span
Calibration

Places indicator into the span calibration routine. Scrolling down with
the #2 key one level begins the procedure.

Press #2 key to
begin sequence

F18
View Calibra-
tion

Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the #2 key one level begins the procedure.

Press #2 key to
begin sequence

F19
Key-in Zero

Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the #2 key one level begins the
procedure.

Press #2 key to
begin sequence

F20
Key-in Span

Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the #2 key one level begins
the procedure.

Press #2 key to
begin sequence

F21
Factory Reset

This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!

Press the #2 key
twice to execute.

SUB-MENU

TITLE

SELECTIONS

F1

Graduations

6,000 to 50,000

F3

Zero Tracking
Band (SAZSM)

0d
5d

F4

Zero Reset
Range

100% (Canada
Only)

F5

Motion Band

3d
5d
10d

F6

Digital Filter

1
2
4

Table 4-1: Invalid Setup Menu selections for commercial applications