4-2

F7

Overload Limit

Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary unit
selected in F8.

"FS" = Full scale in primary units.

FS
FS + 2%

FS + 1d
FS + 9d

F8

Calib. Unit

Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.

1

2

F9

Display
Divisions

Determines the desired weight increments. Value should be
consistent with legal requirements.

1

2
5

F10
Decimal Pt.

Determines location of the decimal point.

0

√ 0.0

0.00 0.000
0.0000 00

F16

Zero
Calibration

Places indicator into the zero calibration routine. Scrolling down with
the ZERO key one level begins the procedure.

Press ZERO key
to begin sequence

F17

Span
Calibration

Places indicator into the span calibration routine. Scrolling down with
the ZERO key one level begins the procedure.

Press ZERO key
to begin sequence

F18

View Calibra-
tion

Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the ZERO key one level begins the procedure.

Press ZERO key
to begin sequence

F19

Key-in Zero

Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the ZERO key one level begins
the procedure.

Press ZERO key
to begin sequence

F20

Key-in Span

Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the ZERO key one level
begins the procedure.

Press ZERO key
to begin sequence

F21

Factory Reset

This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!

Press the ZERO
key twice to
execute.

200E and 200ES only

4

NAME/CODE DESCRIPTION CODE/VALUE

F24

Piece Count
Enable

Used to enable or disable the piece count feature. This feature must
be disabled for commercial applications.
"0" = Disable piece count "1" = Enable piece count

0

1

4

Won’t appear on older units.