Taylor-Wharton 24K w/ AutoTend User Manual
Page 42

above the STOP FILL thermistor assignment. For this reason START FILL cannot be
assigned to thermistor number 1. Similarly, STOP FILL cannot be assigned to ther-
mistor number 7. The factory installs this seven thermistor assembly all the way to the
floor of the sensor tube, which positions thermistor number 1 one inch above the floor
(i.e.; the first thermistor is offset from the floor zero inches).
1.
If you would like to stop the fill at a depth greater than seven inches from the
floor, the sensor assembly must physically be raised inside the sensortube. The
distance that it has been raised becomes the new OFFSET and the controller
must be notified of this offset from the floor. The factory settings are START FILL
at 3 inches and STOP FILL at 6 inches, with an OFFSET of zero. If you want to
establish a START at 15 and STOP at 18, raise the sensor assembly 12 inches
(18 minus 6) and set the controller OFFSET to 12 by: Selecting MENU, LEVEL
SENSING, SET SENSOR OFFSET. The unit should begin filling. The EX LOW
LEVEL sensor has been subjected to the gaseous condition and should sound an
alarm (O.K. to MUTE). Note that in this scenario, you have directed your LN
2
supply to accomplish up to a 15 inch fill. It may be necessary to stop the fill
manually and allow the supply vessel time to regain enough pressure to complete
this exceptionally long fill.
Power Failure Alarm
If power is interrupted, a note of the time and duration will be recorded in the log. No
local alarm will be sounded unless a high temperature is detected after power re-
sumes. The Remote Alarm will be triggered anytime the power is interrupted.
Factory Settings
To restore factory settings at any time KRYOS can be re-initialized. Please note that
a re-initialization clears all logs and passwords and resets settings to those listed in
the table below. To reset factory settings in the control, go to the menu choice
ABOUT THIS CONTROL. (MENU, 6,4) and while looking at this screen type in 1973.
The control will lead you through the process required to reset the factory settings.
Temperature:
Thermocouple #1
Enabled
Thermocouple #2
Disabled
Hi Temp #2 Alarm
-100º C
Hi Temp #2 Alarm
-100ºC
Level Control:
OFFSET
0 inches
START FILL Sensor
3 inches
STOP FILL Sensor
6 inches
Sensor Assembly
Freeze-Guard
System:
RS-232
Enabled
Handshaking
CTS/RTS
Baud Rate
19200
Date & Time
USA Central Time
Brightness
100%
Defog Timer
15 seconds
Lid Switch
Enabled
Auto Defog Timer
15 seconds
Quick-chill
15 seconds
Temperature Control:
Temperature Control
Disabled
Temperature
-150ºC
Temperature Range
+2ºC
Alarms:
LN
2
Supply Alarm
30 minutes
Sensor Error
Enabled
Remote Alarm
30 minutes
Lid Open Too Long
10 minutes
Thermocouple Open Alarm
Enabled
Logging:
Sensor Error Log
Enabled
Low LN
2
Supply Log
Enabled
Remote Alarm Log
Enabled
Open Thermocouple
Enabled
Hi Temp #1 Log
Enabled
Hi Temp #2 Log
Enabled
Fill Logging
Enabled
Lid Open/Closed Log
Enabled
User Access Log
Disabled
Thermocouple #1 Rate Log
Disabled
Thermocouple #2 Rate Log
Disabled
Freeze-Guard:
Valve Open Duration
Disabled
Valve De-icing
Enabled