Taylor-Wharton 24K w/ AutoTend User Manual

Page 42

background image

above the STOP FILL thermistor assignment. For this reason START FILL cannot be
assigned to thermistor number 1. Similarly, STOP FILL cannot be assigned to ther-
mistor number 7. The factory installs this seven thermistor assembly all the way to the
floor of the sensor tube, which positions thermistor number 1 one inch above the floor
(i.e.; the first thermistor is offset from the floor zero inches).

1.

If you would like to stop the fill at a depth greater than seven inches from the
floor, the sensor assembly must physically be raised inside the sensortube. The
distance that it has been raised becomes the new OFFSET and the controller
must be notified of this offset from the floor. The factory settings are START FILL
at 3 inches and STOP FILL at 6 inches, with an OFFSET of zero. If you want to
establish a START at 15 and STOP at 18, raise the sensor assembly 12 inches
(18 minus 6) and set the controller OFFSET to 12 by: Selecting MENU, LEVEL
SENSING, SET SENSOR OFFSET. The unit should begin filling. The EX LOW
LEVEL sensor has been subjected to the gaseous condition and should sound an
alarm (O.K. to MUTE). Note that in this scenario, you have directed your LN

2

supply to accomplish up to a 15 inch fill. It may be necessary to stop the fill
manually and allow the supply vessel time to regain enough pressure to complete
this exceptionally long fill.

Power Failure Alarm
If power is interrupted, a note of the time and duration will be recorded in the log. No
local alarm will be sounded unless a high temperature is detected after power re-
sumes. The Remote Alarm will be triggered anytime the power is interrupted.

Factory Settings
To restore factory settings at any time KRYOS can be re-initialized. Please note that
a re-initialization clears all logs and passwords and resets settings to those listed in
the table below. To reset factory settings in the control, go to the menu choice
ABOUT THIS CONTROL. (MENU, 6,4) and while looking at this screen type in 1973.
The control will lead you through the process required to reset the factory settings.

Temperature:
Thermocouple #1

Enabled

Thermocouple #2

Disabled

Hi Temp #2 Alarm

-100º C

Hi Temp #2 Alarm

-100ºC

Level Control:
OFFSET

0 inches

START FILL Sensor

3 inches

STOP FILL Sensor

6 inches

Sensor Assembly

Freeze-Guard

System:

RS-232

Enabled

Handshaking

CTS/RTS

Baud Rate

19200

Date & Time

USA Central Time

Brightness

100%

Defog Timer

15 seconds

Lid Switch

Enabled

Auto Defog Timer

15 seconds

Quick-chill

15 seconds

Temperature Control:

Temperature Control

Disabled

Temperature

-150ºC

Temperature Range

+2ºC

Alarms:

LN

2

Supply Alarm

30 minutes

Sensor Error

Enabled

Remote Alarm

30 minutes

Lid Open Too Long

10 minutes

Thermocouple Open Alarm

Enabled

Logging:

Sensor Error Log

Enabled

Low LN

2

Supply Log

Enabled

Remote Alarm Log

Enabled

Open Thermocouple

Enabled

Hi Temp #1 Log

Enabled

Hi Temp #2 Log

Enabled

Fill Logging

Enabled

Lid Open/Closed Log

Enabled

User Access Log

Disabled

Thermocouple #1 Rate Log

Disabled

Thermocouple #2 Rate Log

Disabled

Freeze-Guard:

Valve Open Duration

Disabled

Valve De-icing

Enabled