No.
Description
_______________________________________________________________________
1101
Message:
"No scan modes found in nb_code ..."
1102
"Something wrong with byteorder in nb_code ..."
1105
"No suitable scanning mode found in nb_code".
Reason:
File $MARTABLEDIR/mar2300.XXX and/or
$MARTABLEDIR/mar3450.XXX may be corrupted or empty.
Action:
Check file sizes (73 MB and 103MB). Use command:
catmar $MARTABLEDIR/mar2300.XXX
to look at the calibration file header.
1103
Message:
"Scanner serial number in nb_code differs from config"
Reason:
The scanner no. in the file headers of files
$MARTABLEDIR/mar2300.XXX and/or
$MARTABLEDIR/mar3450.XXX are not identical to
$MAR_SCANNER_NO. The calibration files may not
belong to the scanner.
Action:
Call.
1110
Message:
"Cannot create image file"
1111
"Cannot open image file"
1112
"Error writing image array"
1115
"Error writing image header"
Reason:
Insufficient disk space or invalid file permissions.
Action:
Clean up disk or check data directory permissions.
1215/1241
Message:
"Cannot OPEN SHUTTER at START of EXPOSURE"
1234
"SHUTTER was already OPEN at START of EXPOSURE!"
1238
"Cannot CLOSE SHUTTER at END of EXPOSURE!"
Reason:
X−ray beam shutter damaged or dirty. Otherwise, the manual
shutter operation switch mounted on the base may have
been left in the "Open" state.
Action:
Check shutter and manual shutter switch on the base.
Open and close the shutter manually. Is it very hard
to operate?
1239
Message:
"STEPPER error during EXPOSURE!"
Reason:
Problem with PHI axis movement. Possibly software related.
Action:
Mail mar.spy file. Call.