mar345 Program Manual

1 9

______________________________________________________________________________________

3.8 Starting Data Collection

Data collection can be started by pressing the "Go" or "Go/Erase first" button in the
mar345−Change Parameters window. It is a good idea to always let the plate clean first
before actually starting the data collection.

In Multiple Data Sets mode, before actually sending the command to the scanner the

mar345−Run Parameters window will pop up. This window gives an overview ofthe currently
programmed data sets. The buttons in the lower part of the window come with different
layouts:

_______________________________________________________________________

Mode

Left

Center

Right

________________________________________________________________________

Before data

"Go"

"Change"

"Close"

collection

really start

reedit parameters

Close window

During data

"Abort NOW"

"Stop AFTER IMAGE"

Close

collection

Aborts exposure

Stops after current image Close window

________________________________________________________________________

To start data collection you will have to press the "Go" button in this window. During data
collection the mar345−Run Parameters window monitors the progress of the data collection.
The window can be opened by selecting the "Progress" option in the "Windows" submenu
(see 3.2.1). Active data sets are highlighted. Start

−ϕ

and Image will be updated. The number

of oscillations is given by the multiplier in the

∆ ϕ

field. If the scanner is idle, this window

cannot be obtained.

3.9 Stopping Data Collection

Data collection can be stopped by pressing the "Stop" button in the main window or by
pressing the "Stop" or "Abort" button in the mar345−Run Parameters window. If the
scanner is scanning or erasing the imaging plate, this cycle will always be finished. Pressing
"Abort" in this moment will be the same as pressing "Stop". Otherwise, "Abort" closes the
X−ray beam shutter immediately and stops

ϕ

or distance movement.