QBC Star System Operator’s Manual

6000-300-000 7–12

603

Sample door open during rotor
motion error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. If the lid is closed, open it manually (Ref. page 4-10)

a. If the lid opens, turn Instrument Power (Ref. page 4-2)
b. If the error persists contact QBC Technical Support
c. If the lid does not open, contact QBC Technical Support

5. If the lid opens, turn the instrument Power on (Ref. page 4-2)
6. If the error persists contact QBC Technical Support

604

Low internal memory error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power on (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

606

QC label not in the camera field of
view error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. If the lid is closed, open lid manually (Ref. page 4-10)
5. Check and clean the QC label (Ref. page 6-1)
6. Turn the instrument Power on (Ref. page 4-2)
7. If the error persists contact QBC Technical Support

E7 – FSM State

Error Number

Description

Corrective Action

700

General finite state machine error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power on (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

701

Power On Self-Test (POST) general
error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power on (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

702

Power On Self-Test (POST)
software error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power on (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

703

Actuator test failed during Power
On Self-Test (POST)

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Place the instrument into Shipping Prep (Ref. page 2-9)
5. Confirm the rotor spins freely by hand
6. If the rotor spins freely

a. Power down the instrument (Ref. page 4-10)
b. Turn the instrument Power on (Ref. page 4-2)
c. If the error persists contact QBC Technical Support

7. If the rotor does not spin freely, contact QBC Technical Support

704

Encoder failed to initialize during
Power On Self-Test (POST)

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power on (Ref. page 4-2)
5. If the error persists contact QBC Technical Support