6000-300-000 7–9

7 – Troubleshooting

408

Internal printer hardware error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Pull the printer paper release lever towards the front of unit

(Ref. page 3-3)

4. Manually advance the paper using the manual advance wheel

(Ref. page 3-3)

5. If the paper advances, push the printer paper release lever to

lock the paper into place (Ref. page 3-3)

6. Power down the instrument (Ref. page 4-10)
7. Turn the instrument Power on (Ref. page 4-2)
8. If the error persists contact QBC Technical Support

409

Internal printer connection error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Pull the printer paper release lever towards the front of unit

(Ref. page 3-3)

4. Manually advance the paper using the manual advance wheel

(Ref. page 3-3)

5. If the paper advances, push the printer paper release lever to

lock the paper into place (Ref. page 3-3)

6. Power down the instrument (Ref. page 4-10)
7. Turn the instrument Power on (Ref. page 4-2)
8. If the error persists contact QBC Technical Support

410

Internal printer communication
port error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power on (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

411 thru 414

I/O board communication error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power on (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

415

Lid latch micro-switch error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. If the lid is closed, open lid manually (Ref. page 4-10)

a. If the lid opens, turn the instrument Power on (Ref. page 4-2)
b. If the error persists contact QBC Technical Support
c. If the lid does not open, contact QBC Technical Support

5. If the lid opens, turn the instrument Power on (Ref. page 4-2)

416

Lid latch solenoid over current
error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. If the lid is closed, open lid manually (Ref. page 4-10)

a. If the lid opens, turn the instrument Power on (Ref. page 4-2)
b. If the error persists contact QBC Technical Support
c. If the lid does not open, contact QBC Technical Support

5. If the lid opens, turn the instrument Power on (Ref. page 4-2)