beautypg.com

3 sample considerations, 4 sample elution, 5 column calibration – Bio-Rad ENrich™ High-Resolution Size Exclusion Columns User Manual

Page 8

background image

4 ENrich SEC Instruction Manual

2.3 Sample considerations

1. Always filter (0.22–0.45 μm filter) and/or centrifuge your

sample to remove any particulates.

2. In size exclusion chromatography, mass loading is less critical

than the volume of the sample. As a general rule, to obtain
maximum performance, the sample volume should not
exceed 1% of the bed volume (that is, ~0.25 ml for a 24 ml
CV) and the sample concentration should not exceed 20 mg
protein/ml. Larger volumes can be applied, but resolution may
decrease.

2.4 Sample Elution

All species within the sample should elute within 1 full CV. The flow
rate for optimal resolution depends on sample composition and
purity. In general, lower flow rates provide greater performance.
Optimal flow rates for both ENrich SEC 70 and ENrich SEC 650
columns are 0.5–1.25 ml/min (see Table 1).

2.5 Column Calibration

Use the Bio-Rad size exclusion standards (catalog #151-1901).
One vial contains 18 mg of a lyophilized mixture of thyroglobulin
(Mr 670,000), bovine

γ-globulin (Mr 158,000), chicken ovalbumin

(Mr 44,000), equine myoglobin (Mr 17,000), and vitamin B12
(Mr 1,350).