beautypg.com

3 column cleaning, 4 frit replacement – Bio-Rad ENrich™ High-Resolution Size Exclusion Columns User Manual

Page 11

background image

ENrich SEC Instruction Manual 7

3.3 Column Cleaning

Careful preparation (especially filtration) of the sample and the

buffers will maintain the column’s performance and extend

its lifetime. Normally, washing with a few CV of running buffer

with an ionic strength of at least 100 mM NaCl will remove

most contaminants. However, if there is a decrease in column

performance (that is, increasing backpressures or a drop in

resolution), then steps 1 and 2 of the cleaning protocol described

below should be used.
Always reverse the flow during this procedure so any substances

that may be trapped at the top of the column are quickly removed.

During this operation do not exceed more than 50% of the

recommended maximum flow rate (see Table 1). Ensure that the

backpressure stays below the maximum operating pressure.

1. Wash with 2 x 1 ml injections of 2.0 M NaCl followed by 1 CV

of buffer.

2. Wash with 2 x 1 ml injections of 1.0 M NaOH followed by 2 CV

of buffer.

If the column performance has still not returned, the additional

steps below can be followed:

3. Wash with 2 x 1 ml injections of 50% acetic acid followed by

2 CV of buffer.

4. If lipid contamination is suspected, first wash with at least

1 CV of deionized water. Then wash with 1 CV of 70% ethanol

followed by 2 CV of water. The flow rate may need to be

slowed significantly during the 70% ethanol wash.

Before use, re-equilibrate with at least 2 CV of equilibration buffer.

3.4 Frit Replacement

The top frit may need to be replaced after extensive column use

or if increasing backpressures are noticed. Always try to clean the

column in the reverse direction (as described in section 3.3) before

replacing the frit. A frit kit containing a frit removal tool, 2 O-rings,

and 2 frits is available.

Figure 2 shows a column diagram to assist in the replacement of

the top frit.