2 ENrich SEC Instruction Manual

The ENrich support is compatible with aqueous solutions of 6 M
guanidine-HCl and 8 M urea. Detergents and organic solvents
such as methanol, ethanol, and isopropanol may also be used.
In some cases, alcohols and stability agents such as glycerol
may increase the column backpressure. Care should be taken
to maintain a flow rate at which the backpressure is below the
maximum operating pressure.

Table 1. ENrich SEC media and column characteristics.

SEC 70

SEC 650

Linear separation range, Da

500–70, 000

500–650, 000

Column dimensions,
diameter x height, mm

10 x 300

10 x 300

Column volume, ml

24

24

Nominal particle size, µm

10 ± 2

10 ± 2

Recommended flow rates, ml/min*

0.5–1.0

0.75–1.25

Maximum recommended flow rates,
ml/min*

1.5

2.0

Maximum operating pressure

600 psi,

4.1 MPa, 41 bar

600 psi,

4.1 MPa, 41 bar

Recommended sample volume, µl

<250

<250

Efficiency, plates/m

>20,000

>20,000

Working pH range

2–12

2–12

Operating temperatures, °C

4–40

4–40

* At room temperature. Viscosity may increase at lower temperatures, which will reduce the

recommended flow rates.

1.3 Connection to the NGC

and Other

Chromatography Systems

The ENrich columns are fitted with 10-32 type female fittings on either
end. Standard 10-32 fittings can be used to plumb the column to a
Bio-Rad

®

NGC

chromatography system or another vendor’s systems.

Adaptors are available for connection to systems that use 1/4-28 fittings
(catalog # 750-0564).