beautypg.com

1 regular maintenance, 2 bed height adjustment – Bio-Rad ENrich™ High-Resolution Size Exclusion Columns User Manual

Page 10

background image

6 ENrich SEC Instruction Manual

Section 3

: Care of the ENrich SEC

Column

3.1 Regular Maintenance

To maintain performance, the column should be stored in 20%

ethanol, as described below, for storage periods longer than

2 days. Always run at least one CV of water between 20% ethanol

and running buffers.

If the column is exposed to air or if air bubbles are trapped in

the column.

1. Wash the column in the forward direction with 30 ml of water

at 0.25–0.5 ml/min.

2. Wash the column with 30 ml of 20% ethanol at

0.25–0.5 ml/min. Note that the backpressure will increase in

ethanol. Do not exceed 600 psi (4.1 MPa).

3. Wash the column with 30 ml of deionized water at the usual

operating flow rate as recommended in Table 1.

4. Re-equilibrate the column with the desired buffer.

If the column begins to show abnormally high backpressure.

If the backpressure increases abnormally, it is likely that the

column needs to be cleaned and/or the frit at the top of the bed

is clogged. First, clean the column using the method described in

section 3.3. If this is not sufficient, replace the top frit as described

in section 3.4.

3.2 Bed Height Adjustment

Under certain conditions of buffer composition, high flow rates,

or long-term use, the resin bed may compress, creating a void

between the frit and the top of the bed. Normally, the void can

be eliminated by turning the adjusting nut clockwise until the frit

just touches the top of the bed. If the bed compresses at high

flow rates, stop the pump and loosen the top fitting, then use

the adjusting nut to remove the void, retighten the top fitting, and

resume pumping buffer. If the bed has compressed from long-term

use, replace the top frit as a precaution.