12 ENrich Instruction Manual

13. Lower the adaptor to the top of the bed by turning the

adjusting nut clockwise (Figure 8).

Fig. 8. Lower the adaptor by turning the adjusting nut clockwise.

14. Tighten the lock nut by turning it counterclockwise (Figure 9).

Fig. 9. Hand-tighten the lock nut by turning it counterclockwise.

3.4 Storage Conditions

Prior to long-term storage, the column should be cleaned as

previously described and then washed with 3 column volumes of

20% ethanol at a flow rate that is 50% of the recommended flow

rate (see Table 1). This will prevent microbial growth. Store the

column in a safe place at room temperature or 4°C. Never allow

the column to freeze.