background image

10 ENrich Instruction Manual

4. Raise the adaptor a few millimeters by slowly by turning the

adjusting nut counterclockwise (Figure 5).

Fig. 5. Raise the adaptor by turning the adjusting nut counterclockwise.

5. Unscrew the upper retainer. Let it rest on the bottom retainer

(Figure 6).

Fig. 6. Remove the retainer.

6. Slowly pull out the upper adaptor from the glass column.

Allow the buffer flow to continue in order to maintain the

integrity of the top of the bed.