beautypg.com

1 column cleaning – Bio-Rad ENrich™ High-Resolution Ion Exchange Columns User Manual

Page 11

background image

ENrich Instruction Manual 7

Section 3

: Care of the ENrich Ion

Exchange Column

3.1 Column Cleaning

Careful preparation (especially filtration) of the sample and

the buffers will maintain the column performance and lifetime.

Normally, washing with 1.0 M NaCl or KCl will remove most

bound components. However, if there is a significant decrease in

column performance (increasing backpressures or a significant

drop in resolution), then a more extensive cleaning protocol such

as that described in steps 1–6 should be used. Always reverse

the flow during this procedure so tightly bound substances

at the top of the column are removed quickly.

During this operation do not exceed more than 50% of the

recommended flowrates (see Table 1).

1. Wash with 2 column volumes of 2.0 M NaCl or KCl followed

by 3 column volumes of water.

2. Wash with two 100 μl injections of 1.0 M NaOH followed by

3 column volumes of water.

3. Wash with two 100 μl injections of 50% acetic acid followed

by 3 column volumes of water.

4. If lipid contamination is suspected, wash with 1 column

volume of 20% ethanol followed by 3 column volumes of

water.

5. Wash with 2 column volumes of 2.0 M NaCl or KCl, or the

salt containing the desired counter-ion.

6. Wash with 3 column volumes of equilibration buffer.