Bio-Rad Model EG-1 Econo™ Gradient Monitor User Manual
Page 15
12
Larger or smaller than
Chart recorder input
Check chart recorder
expected gradient
voltage set incorrectly
settings (See Sections
profile
4.2 and 5)
If using Gradient
Re-calibrate the
mode, buffer A and
gradient mode
B values not set
correctly
Ghost peaks appear in
Charged sample
No problem
gradient profile
detected (e.g. protein)
Air bubbles passing
Use degassed buffers.
through flow cell
Check for loose plumb-
ing connections
Chart recorder
Chart recorder is not
Check connections (See
not responding
connected properly
Sections 3.2 and 4.2)
Full scale of recorder
Set recorder range to
set too high
1 V
Recorder has bottomed Readjust pen position
out
Flow cell not
Check flow cell
connected in-line
plumbing
No power
Faulty power
Check power cable
connection
connection
Blown fuse
Check fuses (See
Appendix A)
Unit not switched on
Check power on/off
switch on rear panel
Conductivity
Dirty flow cell
Clean flow cell (See
measurements with
Section 6)
same buffer appear
to decrease over time
Problem
Possible Cause
Solution
M7318150H 2/27/1 2:06 PM Page 12