Bio-Rad BioLogic LP System User Manual
Page 11

a.
An SV-3 solenoid valve is used with the Model 2110 fraction
collector.
b.
Plug the SV-3 valve directly into the diverter valve receptacle on
the back of the BioLogic LP instrument.
Section 9
Power Connections
1. Plug the line power cords into the rear panel of the BioLogic LP and
the fraction collector.
2. Power for the 1327 chart recorder is supplied from the 12 volt outlet on
the rear panel of the BioLogic LP (next to the power connection for the
UV optics module). Locate the thin black power cable, and connect
this between the BioLogic LP and the power inlet for the 1327 (next to
the DIN socket, chart recorder rear panel). The 1327 chart recorder
may also be powered by a separate AC adaptor.
Section 10
Pump Setup
1. Adjust platen pressure screw (on pump head) as shown in Figure 4.
2. Using a screwdriver or coin, turn the screw counterclockwise as far as
it will go, then turn clockwise three full turns.
Fig. 4.
1
Alarm
2
3
4
5
6
7
8
0
9
C
.
Platen Adjust
Screw
Orange Lock-Ring
1.6 mm ID Tubing
Female
Luer Fitting
See Detail
7
4006073D.qxp 8/17/2009 7:45 AM Page 7
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- iMark Microplate Absorbance Reader (112 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)