Prepare the standard dilution series – Bio-Rad Bio-Plex Pro™ Human Isotyping Assays User Manual
Page 15
13
Prepare the Standard Dilution Series
The following procedure produces an eight-point standard curve with a
fourfold dilution between each point. Pipet carefully using calibrated pipets
and use a new pipet tip for every volume transfer.
1. Label eight 1.5 ml microcentrifuge tubes S2 through S8 and Blank.
Alternatively, using Titertube
®
micro test tubes may prove to be more
convenient if a multichannel pipet will be used to load the plate.
2. Add 150 µl of the appropriate diluent to tubes S2–S8 and Blank.
(Figure 3).
3. Vortex reconstituted standards at medium speed for 5 sec before
removing any volume. Transfer 50 µl to the S2 tube containing the
chosen standard diluent. Vortex for 5 sec.
4. Use a new pipet tip to transfer 50 µl from the S2 tube to the
S3
tube.
Vortex for 5 sec.
5. Continue with 1:4 (fourfold) serial dilutions as shown in Figure 3.
6. Use reconstituted and diluted standards and controls immediately. Do
not freeze for future use.
Fig. 3. Preparing a fourfold dilution series with a single reconstituted standard.
150 150 150 150 150 150 150 150
S1
S2 S3 S4 S5 S6 S7 S8
Blank
Diluent, µl
50
50 50 50 50 50 50
Transfer Volume, µl
Reconstituted
Standard