Protocol Acceptance by Customer and List of Tests Performed
Bio-Rad Laboratories recommends that the IQ/OQ protocols be performed in total when the
Experion automated electrophoresis station and software are initially installed. Installation
qualification should also be performed when the Experion electrophoresis station is moved
to a new location, when the software is upgraded, and when the computer that runs the
electrophoresis station is changed or modified. Operational qualification should also be run on a
regular basis to confirm that the system is performing to specifications, and also when there is a
question that the instrument is not performing to specifications.
I have reviewed the IQ/OQ document and agree that it provides the appropriate procedures for the
Installation Qualification/Operational Qualification of the Experion automated electrophoresis station
and software.
________________________
_________________________
_________________
Customer Name (print)
Signature
Date
__________________________
___________________________
___________________
Reviewer Name/Title (print)
Signature
Date