Table viii: software oq test – Bio-Rad Experion DNA Analysis Kits User Manual
Page 29
Technical Support: 1-800-424-6723 • www.bio-rad.com
21
Operational Qualification
Experion Electrophoresis Station – Serial Number______
Table VIII: Software OQ Test
Item
Description
Performed by/date
1
Double-click the Experion Software icon on the computer desktop
to launch the software.
2
Select Perform Software OQ from the Validation menu.
3
Select Start OQ. If any part of the test fails, uninstall the software,
reinstall, and repeat first the Software IQ test and then the OQ test.
If there is a repeated failure, call Bio-Rad Technical Support. Details
can only be viewed when there is a failure.
4
Select Next to view details. Select Print to print the report on the
next page. Security Edition only: If further IQ/OQ tests will be per-
formed, print all test results together through Report IQ/OQ Results.
5
Append printed report to IQ/OQ records binder.
Comments:_________________________________________________________________________
Performed by:_________________________________ Date:_______________________________
Reviewed by:__________________________________ Date:_______________________________