beautypg.com

3 specifications, Chef-dr iii specifications – Bio-Rad CHEF-DR® III Variable Angle System User Manual

Page 5

background image

tials are readjusted immediately to maintain uniform fields, thus insuring high resolution. In
PACE, the voltage potential of each of the 24 electrodes is regulated independently. Unlike
the CHEF-DR II system, which has a fixed reorientation (field) angle of 120° due the hexag-
onal geometry of the electrode array, the CHEF-DR III system can generate field angles from
90–120°. In addition, the CHEF-DR III system permits up to three consecutively executing
blocks of run conditions with battery backed-up RAM and automatic restart after power
interruption.

1.3 Specifications

These specifications pertain to the complete CHEF-DR III system. A system including the

Cooling Module is also available.

CHEF-DR III Specifications

Power Module:

Dimensions

43 (depth) x 48 (width) x 17.5 (height) cm

Construction

Aluminum chassis

Weight

10 kg

Power supply

350 V maximum, to allow maximum gradient of
9 V/cm, continuously adjustable; built in

Electrical

Maximum amps

0.5 amperes

Allowable voltage gradients

0.6–9 V/cm, in 0.1 V/cm increments

Battery back-up

All parameters in memory

Electrode potentials

Dynamically regulated (feedback adjustment)
+/- 0.5%

Data entry

Keyboard

Functional

Switching range

0.1 sec to 65K sec

Switch angle variable

90–120 degrees (all electronic switching) in 1°
increments

Maximum program blocks

3, with automatic execution

Maximum run time

999 hours per block

Input voltage range

100–120 VAC/50–60 Hz/4 amps
220–240 VAC/50–60 Hz/2 amps

Fuses

0.5 amp Fast Blow for high voltage output
3.15 amp Slow-Blow (100/120 V) or
1.60 amp Slow-Blow (220/240 V)

Environmental

Operating

50 °F (10 °C) to 90 °F (32 °C) temperature
30–80% humidity

Storage

32 °F (0 °C) to 140 °F (60 °C) temperature
10-90% humidity

Electrophoresis cell:

Dimensions

11.4 x 44.2 x 50.3 cm, horizontal format

Construction

Cover: Vacuum formed polycarbonate

Base: Injection molded polycarbonate

Lid

Safety interlocked

Weight 10.2

kg

Electrodes

24, platinum (0.02 inch diameter)

Temperature monitoring

Via precision temperature probe mounted in base
of cell

2