3 specifications, Chef-dr iii specifications – Bio-Rad CHEF-DR® III Variable Angle System User Manual
Page 5
tials are readjusted immediately to maintain uniform fields, thus insuring high resolution. In
PACE, the voltage potential of each of the 24 electrodes is regulated independently. Unlike
the CHEF-DR II system, which has a fixed reorientation (field) angle of 120° due the hexag-
onal geometry of the electrode array, the CHEF-DR III system can generate field angles from
90–120°. In addition, the CHEF-DR III system permits up to three consecutively executing
blocks of run conditions with battery backed-up RAM and automatic restart after power
interruption.
1.3 Specifications
These specifications pertain to the complete CHEF-DR III system. A system including the
Cooling Module is also available.
CHEF-DR III Specifications
Power Module:
Dimensions
43 (depth) x 48 (width) x 17.5 (height) cm
Construction
Aluminum chassis
Weight
10 kg
Power supply
350 V maximum, to allow maximum gradient of
9 V/cm, continuously adjustable; built in
Electrical
Maximum amps
0.5 amperes
Allowable voltage gradients
0.6–9 V/cm, in 0.1 V/cm increments
Battery back-up
All parameters in memory
Electrode potentials
Dynamically regulated (feedback adjustment)
+/- 0.5%
Data entry
Keyboard
Functional
Switching range
0.1 sec to 65K sec
Switch angle variable
90–120 degrees (all electronic switching) in 1°
increments
Maximum program blocks
3, with automatic execution
Maximum run time
999 hours per block
Input voltage range
100–120 VAC/50–60 Hz/4 amps
220–240 VAC/50–60 Hz/2 amps
Fuses
0.5 amp Fast Blow for high voltage output
3.15 amp Slow-Blow (100/120 V) or
1.60 amp Slow-Blow (220/240 V)
Environmental
Operating
50 °F (10 °C) to 90 °F (32 °C) temperature
30–80% humidity
Storage
32 °F (0 °C) to 140 °F (60 °C) temperature
10-90% humidity
Electrophoresis cell:
Dimensions
11.4 x 44.2 x 50.3 cm, horizontal format
Construction
Cover: Vacuum formed polycarbonate
Base: Injection molded polycarbonate
Lid
Safety interlocked
Weight 10.2
kg
Electrodes
24, platinum (0.02 inch diameter)
Temperature monitoring
Via precision temperature probe mounted in base
of cell
2