beautypg.com

Bio-Rad CHEF-DR® III Variable Angle System User Manual

Page 31

background image

Section 7
Troubleshooting Guide

Problem

Solution

Equipment

No power

1. Check fuse at back of power module
2. Check source of A/C power
3. Contact Bio-Rad Laboratories

No voltage across electrodes,

1. Check that lid is on

with AC power light on

2. Check that serial cable is fully inserted at both ends
3. Check that coiled interlock cable is firmly attached
4. Confirm that all leads are properly attached
5. Check HV fuse on front panel (0.5 A FB) and replace

if necessary

6. Check that the unit is not set in pause mode

Gel floats away

1. Pump flow rate is too high. Adjust with Variable

Speed Pump.

No or low buffer flow

1. Look for kink in tubing
2. Check Cooling Module; buffer in the heat exchanger

can freeze if the chiller is cooling, but the pump is not on

3. Check pump connection and if the pump is on
4. Electrophoresis chamber outlet port may be clogged.

Reverse buffer recirculation flow to unclog the ports.
Remove agarose debris from the chamber and restore
the flow to original condition.

Gel band patterns appear very

1. Foreign object in chamber (remove thermometer, etc.)

distorted; lanes very curved,

*

2. Insufficient or non-uniform cooling due to low pump

bands sharp, but slanting

flow

3. Check that buffer is level with surface of gel. Use

leveling feet.

4. Check that the current is equal for both switch direc-

tions

5. Replace damaged electrode
6. Power module fault; contact Bio-Rad Laboratories
7. Not enough buffer in cell (total should be 2.2 liters)
8. Check to insure the electrophoresis chamber is level

(Section 4.2)

* Slight distortion of the outermost lane is normal

Applications

Bands smeary or fuzzy

1. Excessive heating. Lower the voltage or ionic

strength of the buffer.

2. Improper switch interval. See Section 5.1.
3. Gel percentage too low. Increase.
4. Sample degraded. Impure enzymes, or wash cycles

too short (agarose blocks).

5. Agarose impurities. Consult Bio-Rad.
6. Sample overload. Adjust sample concentration.

28