Bio-Rad PROTEAN® IEF Cell Accessories User Manual
Page 15
Problem Cause
Solution
3. Sample wicks
The sample cups are
For optimum performance use new
out of sample
disposable. When re-using
sample cups for each focusing run.
cups.
the sample cups, salts
can build up in the
corners and provide a
capillary path for the
sample to travel.
4. Programmed
Final combined conductivity
Run at maximum attainable voltage for
max. voltage is
of the IPG strip, sample,
longer time to arrive at the required
not reached
and rehydration buffer
number of volt-hours.
with a current
determines the maximum
limit of 50 µA
attainable voltage.
If possible, reduce the conductivity of
per strip.
the rehydration buffer by reducing the
Bio-Lyte
®
concentration.
Increase the current limit per strip.
However, higher current limits generate
more heat and increase the possibility of
burned regions in the IPG strips.
Section 6
Warranty
The Cup Loading Tray is warranted for 1 year against defects in materials and workmanship. If
any defects should occur during this warranty period, Bio-Rad Laboratories will replace the
defective parts without charge. However, the following defects are specifically excluded:
•
Defects caused by improper operation
•
Repairs or modifications performed by anyone other than Bio-Rad Laboratories or their
authorized agent
•
Damage caused by accidental misuse
•
Damage caused by disaster
•
Common replacement parts including platinum wire
•
Damage caused by use of organic solvents or basic detergents
For inquiries or to request repair service, contact your local Bio-Rad office.