Bio-Rad PROTEAN® IEF Cell Accessories User Manual

Page 14

background image

Section 5
Troubleshooting

Problem Cause

Solution

1. Sample is

Cup positioning is incorrect.

When positioning the cups, press down

leaking from

on the connecting bridge until it stops

the sample

on the channel wall of the focusing tray.

cups.

(See illustration on page 7 .)

The cut edge from the

Make sure to use a sharp cutting

connecting bridge contains

device such as scissors to evenly cut

a burr that prevents the

the connecting bridge. Inspect cut to

connecting bridge from

ensure that it does not contain any

proper positioning on the

burrs.

channel wall.

The cup is positioned in an

Make sure the IPG strips are rehydrated

IPG area that is not

evenly and thoroughly. See ReadyStrip

completely rehydrated.

IPG Strip Manual for correct rehydration
directions. Make sure the correct
volume is used for the selected IPG
strip. See table in Section 2.2.

The cup itself is

Replace cup.

malfunctioning.

2. Initial low or

Incorrect positioning of the

Make sure the electrodes are positioned

zero current

electrode assemblies.

directly on the IPG strips. See page 6

(no electrical

for correct positioning.

contact)

When using electrode wicks make sure
they are wetted properly and are placed
on the IPG strips and the electrode
assemblies are in contact with the
electrode wicks.

Make sure the movable electrode
assemblies make contact with the
electrode contact area on the focusing
tray.

Platinum wire on electrode

Inspect the platinum electrode wire to

holder has been

make sure it is intact. If in doubt, use a

compromised.

voltmeter to check for continuity.

12