beautypg.com

Bio-Rad ddPCR™ Supermix for Probes User Manual

Page 17

background image

QX200 Droplet Generator Instruction Manual

| 9

Chapter 2 Droplet Generation

Chapter 2 Droplet Generation

8. When droplet generation is complete, all three indicator lights are solid green. Open the door by pressing

the button, and remove the holder (with DG8 cartridge still in place) from the unit. Remove the disposable
gasket from the holder and discard it. The top wells of the cartridge contain droplets, and the middle and
lower wells are nearly empty with a small amount of residual oil.

QX200 Droplet Generator Instruction Manual

Table 2.2. Status indicator lights on the QX200 droplet generator. If the central LED flashes amber, the gasket is
not placed on the holder correctly or is missing and no seal was made. If the right LED flashes amber, a process error
occurred because the volume is too low in at least one well.

Solid green

Power on

DG8 cartridge holder in place

Run complete

Flashing green

Run in progress

Flashing amber

No seal; no gasket or empty well

Low volume in well

Off

Power off

No DG8 cartridge holder

Idle

QX200 droplet generator with DG8 cartridge in place.

Keep the DG8 cartridge in the holder.

Press to open and close

Status indicator lights