Troubleshooting – LaMotte SMART 3 Electronic Soil Lab User Manual

Page 49

background image

SMART3 Soil 1.11

49

reuse, and recycling systems contact local or regional waste administration or

recycling services.

TROUBLESHOOTING

ERROR MESSAGES

OVER RANGE

If the message OVERRANGE is displayed when scanning a sample, the

sample may be over range or under range. If the sample is over range the

sample should be diluted and tested again (see Sample Dilution Techniques

and Volumetric Measurements, page 12).

If overrange is displayed, press

ENTER

to continue testing on

diluted samples.
NOTE: After pressing

ENTER

, the

overrange concentration will be
displayed. This concentration is
an approximation only.

002 Aluminum

Overrange
continue
Scan Blank

Scan Sample

12:00:00 001/500

CALIBRATION

As with all pre-calibrated meters, it is highly recommended, even if not required

by regulations, that the user periodically verify the performance of the meter

by running standards with a predetermined concentration. Results outside of

specification are an indication that the meter needs to be adjusted. This can be

done following the user calibration described on page 24. If the user calibration

fails to properly adjust the meter then the meter should be returned to LaMotte

Company for recalibration. (See page 48).

STRAY LIGHT

The SMART3 Colorimeter should have no problems with stray light. Make sure

that the sample compartment lid is always fully closed, except when testing

COD with the adapter.