LaMotte Tracer Dissolved Oxygen Meter User Manual
Page 4
INTRODUCTION
Congratulations on your purchase of the Dissolved Oxygen/Temperature
Tracer. Units of measure are % saturation, mg/L or ppm for dissolved
oxygen and °C or °F for temperature. Advanced features include data hold,
25 point memory, auto shut off, automatic temperature compensation, and
salinity/altitude compensation. Careful use and maintenance will provide
years of reliable service.
SPECIFICATIONS
Display
Dual function 3 ½ digit LCD with Bar Graph
Display size: 24 mm x 20 mm
Range
% Sat: 0 to 200.0%
DO: 0 to 20.00 mg/L, 0 to 20.00 ppm
Temp: 0 to 50°C, 32 to 122°F
Resolution:
% Sat: 0.1%
DO: 0.01 mg/L, 0.01 ppm
Temp: 0.1°C, 0.1°F (0 to 99°F),
1.0 °F (>100 °F)
Accuracy:
% Sat: ±2.0% full scale
DO: ±2% full scale
Temp: ±1.0°C, ±1.8°F
Operating Temperature
Range
0 to 50°C (32 to 122°F)
Automatic Temperature
Compensation
0 to 50°C (32 to 122°F)
Salinity Compensation
0 to 50 ppt in 1 ppt increments
Altitude Compensation
0 to 20,000 ft in 1000 ft increments
Sensor
Polarographic type
Membrane
Bonded membrane cap with threaded fitting
Data Storage
25 tagged (numbered) data sets with recall
Auto Power Off
After 10 minutes of no button presses
(override available)
Battery Power
Four 303/357 button batteries
Low Battery Indicator
‘BAT’ appears on LCD
Dimensions
1.4 X 6.8 X 1.6 inches (36 X 173 X 41 mm)
4