beautypg.com

LaMotte Tracer Dissolved Oxygen Meter User Manual

Page 14

background image

Function/Resulting
Action

Power Mode

Button Press
Sequence

Enter Salinity
Compensation

On

Any

Press and release
CAL/RECALL button
twice. Displays SAL

Change Salinity
Compensation

On

SAL

Momentary press of
MODE/HOLD button.
Each press increases by 1
ppt from 0 to 50 ppt.

Exit Salinity
Compensation

On

SAL

Press and release
CAL/RECALL button for
2 seconds to enter
Altitude Compensation.
Or press and release the
CAL/RECALL button
one more time to enter
the measurement mode.

Enter Altitude
Compensation

On

Any or
SAL

Press CAL/Recall button
twice to enter SAL. Press
CAL/RECALLfor 2
seconds to enter Altitude
Compensation mode.
Displays Ald

Change Altitude
Compensation

On

Ald

Momentary press of
MODE/HOLD button
Each press increases by
1000 ft from 0 to 20,000
ft.

Exit Altitude
Compensation

On

Ald

Press CAL/RECALL
button to exit and save
changes.

Change Temperature
Units

Off

N/A

Press and hold
CAL/RECALL button
then momentarily press
ON/OFF button. Release
CAL/RECALL button
when SELF CAL is
displayed.

14