6 technical specifications, 1 technical specifications, Technical specifications – Metrohm 850 Professional IC Detector - iDetector User Manual
Page 25
■■■■■■■■■■■■■■■■■■■■■■
6 Technical specifications
2.850.9010 IC Conductivity Detector
■■■■■■■■
17
6 Technical specifications
6.1
Technical specifications
Type
■
Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)
■
Intelligent detector with 6 reference chromatograms
Measuring range
0…15000 µS/cm without range switching
Noise
< 0.1 nS at 1 µS/cm
Deviations from
the linearity
■
< 0.1 % for conductivity values higher than 16 µS/cm.
■
< 1 % for conductivity values lower than 16 µS/cm.
drift
< 0.2 nS/cm per hour
Measuring rate
10 measurements per second for optimum results without filtering
Resolution
0.0047 nS/cm
Baseline
Noise < 0.2 nS/cm typical for sequential suppression
Conductivity
detector
Cell volume
0.8 µL
Cell constant
■
Individual calibration data saved in the detector
■
adjustable in the range: 13.0…21.0 /cm
Electrodes
Ring-shaped electrodes made from stainless steel
Materials in
contact with
eluent
Chemically inert PCTFE
Maximum oper-
ating pressure
5.0 MPa (50 bar)
Cell tempera-
ture
20…50 °C in increments of 5 °C
Temperature
stability
< 0.001 °C
Temperature
compensation
0…5 %/K adjustable, default 2.3 %/K
Heating up time
< 30 minutes (40 °C)