beautypg.com

5 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 850 Professional IC Detector - iDetector User Manual

Page 24

background image

5.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

2.850.9010 IC Conductivity Detector

5 Troubleshooting

5.1

Problems and their solutions

Problem

Cause

Remedy

Marked rise in pres-
sure

Conductivity detector
blocked.

(see Chapter 4, page 14)

Shorten capillary ends by several mm .

Rinse detector against the normal flow
direction.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the connection of the detector
cable.

Switch the instrument off and on again
(after 15 seconds).