2 Installation

2.850.9010 IC Conductivity Detector

■■■■■■■■

9

NOTE

With the instruments of the 850 Professional line of instruments, up to
two detectors can be placed and connected.

Placing the detector in the instrument

1 Removing the knurled screws

Unscrew knurled screws on the rear panel.

Optional with 850 instruments: Remove the handle if it is still fas-
tened to the instrument.

2 Removing rear panel

Remove rear panel.

3 Positioning the detector

Put the detector through the opening and position it on the sup-
port surface intended for this purpose and slide it as far as the
limit stop to the front.

4 Replacing rear panel

Insert the detector cable in a cable feed-through on the rear
panel.

Place the detector outlet capillary in a suitable capillary feed-
through.

Replacing the rear panel.

The detector cable (3-1) and the detector outlet capillary (3-2)
should freely hang out of the cable or capillary feed-through.

5 Tightening the knurled screws

Optional with 850 instruments: Reattach the handle higher up
and use it as a holder for MPaks.

Tighten the knurled screws.