2.4 Connecting the detector inlet capillary

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

2.850.9010 IC Conductivity Detector

Connecting detector inlet capillary to MSM / MSM-HC

1 Connecting detector input

Connect the detector inlet capillary (5-1) and the capillary of the
MSM / MSM-HC (5-2)
labeled with out to one another using a
6.2744.040 coupling (5-3) and two 6.2744.070 short pressure
screws (5-4)
.

1

2

3

4

Figure 5

Connector detector – MSM

1

Detector inlet capillary

2

MSM / MSM-HC outlet capillary
labeled with out.

3

Coupling 6.2744.040

4

Pressure screws, short 6.2744.070

Connecting the detector inlet capillary to the MCS

1 Connecting detector input

Fasten the detector inlet capillary (6-1) using a 6.2744.090 long
PEEK pressure screw (6-2) on the output of the MCS (6-3)
.

3

2

1

Figure 6

Connection detector – MCS

1

Detector inlet capillary

2

Pressure screw, long 6.2744.090

3

MCS output