beautypg.com

3 connecting the detector, Connecting the detector – Metrohm 850 Professional IC Detector - iDetector User Manual

Page 18

background image

2.3 Connecting the detector

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

2.850.9010 IC Conductivity Detector

2.3

Connecting the detector

Connecting the detector cable and the detector outlet
capillary

CAUTION

The instrument must be switched off when connecting a detector.

1 Connecting the detector

Connect the detector connection cable (3-1) to the detector con-
nection socket Detector 1.

NOTE

Recommendation: In AnCat systems with 2 detectors: anions to
Detector 1, cations to Detector 2.

2 Connecting the detector outlet capillary

Guide the detector outlet capillary (3-2) into a sufficiently large
waste container and fasten it there.

NOTE

The detector outlet capillary must be free of blockages (the mea-
suring cell is tested to 5 MPa = 50 bar backpressure).

NOTE

The detector outlet capillary must not be shortened!

Shortening the detector outlet capillary leads to increased noise!