4.6 Method 4

824 Easy Sample Changer,

Standard methods

33

4.6 Method 4

Use:

For simple titrations (including KF titrations) or pH measurements (with-

out calibration). The initial addition of auxiliary reagents or solvents is

possible.

Required instruments:

Metrohm titrator (Titrino or Titrando) or pH meter

Metrohm Dosimat (optional)

Metrohm 722/802 Rod Stirrer or 741 Magnetic Stirrer

2 external pumps (e.g. Metrohm 772 Pump Unit; Pump 1 for rins-

ing, Pump 2 for aspirating

Features:

After each determination the sample vessel is emptied by Pump 1 and

then the electrode is rinsed by Pump 2. The sample vessel is then emp-

tied again by Pump 1.

Preparation:

Fill a sufficiently large conditioning vessel with a suitable solvent and

place it in the highest rack position.

Procedure:

(Start sequence) Initialize the sample changer

Move to sample and lower lift to working position

Switch on stirrer

If a Dosimat is connected (optional): add auxiliary reagent or sol-

vent and wait until addition is completed

Start titration/measurement and wait until it is completed

Switch off stirrer

Move lift to rinsing position and empty sample vessel

Rinse electrode and empty sample vessel again

Move lift to rotating position and allow electrode to drip

(Final sequence) Move to conditioning beaker and immerse elec-

trode