6 method 4, Method 4 – Metrohm 824 Easy Sample Changer User Manual
Page 39

4.6 Method 4
824 Easy Sample Changer,
Standard methods
33
4.6 Method 4
Use:
For simple titrations (including KF titrations) or pH measurements (with-
out calibration). The initial addition of auxiliary reagents or solvents is
possible.
Required instruments:
•
Metrohm titrator (Titrino or Titrando) or pH meter
•
Metrohm Dosimat (optional)
•
Metrohm 722/802 Rod Stirrer or 741 Magnetic Stirrer
•
2 external pumps (e.g. Metrohm 772 Pump Unit; Pump 1 for rins-
ing, Pump 2 for aspirating
Features:
After each determination the sample vessel is emptied by Pump 1 and
then the electrode is rinsed by Pump 2. The sample vessel is then emp-
tied again by Pump 1.
Preparation:
Fill a sufficiently large conditioning vessel with a suitable solvent and
place it in the highest rack position.
Procedure:
•
(Start sequence) Initialize the sample changer
•
Move to sample and lower lift to working position
•
Switch on stirrer
•
If a Dosimat is connected (optional): add auxiliary reagent or sol-
vent and wait until addition is completed
•
Start titration/measurement and wait until it is completed
•
Switch off stirrer
•
Move lift to rinsing position and empty sample vessel
•
Rinse electrode and empty sample vessel again
•
Move lift to rotating position and allow electrode to drip
•
(Final sequence) Move to conditioning beaker and immerse elec-
trode