beautypg.com

3 technical data, 1 control, 2 curves – Metrohm 822 Titration Curve Simulator User Manual

Page 26: 3 analog output, Technical data, Control, Curves, Analog output

background image

4.3 Technical data

18

822 Titration Curve Simulator, Instructions for Use

4.3 Technical

data

4.3.1 Control

parallel

serial

Remote

8 bit / 25 bit

4.3.2 Curves

Curve memory

Number of curves

Flash

8

4 Mbit

Curve types

Acid-Base

Citric acid

Recorder control

Conditioning

Karl Fischer

Diagonal

D/A converter test

Testing department test

+240 mV ... –300 mV

Endpoint at (50.00 ± 0.03) % of the cylinder volume

+250 mV ... –300 mV

1

st

endpoint at (17.50 ± 0.15) % of the cylinder volume

2

nd

endpoint at (35.0 ± 0.1) % of the cylinder volume

3

rd

endpoint at (52.50 ± 0.02) % of the cylinder volume

steps of 200 mV per 2.5 % of the cylinder volume over

the whole voltage range

series 02, 03: –2.5 V ... +2.5 V ... –2.5 V

from series 04: –2 V ... +2 V ... –2 V

series 02, 03: end at 10 % of the cylinder volume

from series 04: end at 8.4 % of the cylinder volume

Endpoint at (26.15 ± 0.10) % of the cylinder volume

series 02, 03: –2.5 V ... +2.5 V

from series 04: –2 V ... +2 V

–2.5 V ... +2.5 V

–2.0 V ... +2.0 V

Reproducibility of the endpoints (volume)

Acid-Base

Citric Acid

1

st

endpoint

2

nd

endpoint

3

rd

endpoint

Karl Fischer

99.9 %

97.5 %

99.5 %

99.95 %

99.5 %

4.3.3 Analog

output

–2.5 V ... +2.5 V

Range