beautypg.com

4 testing department – test, 2 meaning of the leds, 1 serial 8-bit – Metrohm 822 Titration Curve Simulator User Manual

Page 21: Testing department – test, Meaning of the leds, Serial 8-bit

background image

4

Appendix

822 Titration Curve Simulator, Instructions for Use

13

4.1.4 Testing department – test

A ramp in the voltage range of –2 V to +2 V is run through with curve

"7".

test curve

-2.0

-1.5

-1.0

-0.5

0.0

0.5

1.0

1.5

2.0

0

5000

10000

15000

20000

25000

30000

pulse number

vo

lta

g

e [V

]

Excel diagram from original data

4.2 Meaning of the LEDs

The green LEDs represent the commands (more exactly: the single bits)

from the control device for the dosing device. The red LEDs represent

the reply i.e. the single bits from the dosing device to the control device.

Labeling on the device:

Red LEDs

serial out (SER_OUT)

Green LEDs

serial in (SER_IN)

serial 8-bit

serial 25-bit

bit-No.

serial 25-bit

serial 8-bit

e. g.:

685 Dosimat

700 Dosino

e. g.:

805 Dosimat,

800 Dosino

e. g.:

805 Dosimat,

800 Dosino

e. g.:

685 Dosimat,

700 Dosino

Reply from the dosing device

0 ... 24 Commands for the dosing device

Status information from 822

---

Status information from 822

4.2.1 Serial

8-bit

With the 8-bit transmission the single bits 0...7 have the following mean-

ings:

Reply from the dosing device:
SER_OUT - bit 0:

BUSY

Status message for active cock

switching

SER_OUT - bit 1:

STOP

Status message if difference counter

= zero

SER_OUT - bit 2:

ESU

End switch down: cylinder at minimal

stroke

SER_OUT - bit 3:

EXP0

Bit 0 for coding the cylinder volume