2 connection, 1 connection to the msb interface, Connection – Metrohm 822 Titration Curve Simulator User Manual
Page 11: Connection to the msb interface, Connection to titrandos with internal dosing drive
![background image](/manuals/586974/11/background.png)
2
Connection
822 Titration Curve Simulator, Instructions for Use
3
2 Connection
Titrators with internal dosing drive do not need an additional dosing de-
vice. Titrators without internal dosing drive (e. g. 809/836 Titrando,
726/796
Titroprocessor) need a connected dosing device
(685/805 Dosimat or 700/800 Dosino). The Curve Simulator is con-
nected between titrator and dosing device.
A detailed list of the connecting combinations can be found in the draw-
ings given below.
Note!
The 822 Titration Curve Simulator does not replace a dosing device!
The communication protocol 1 has to be adjusted depending on the
type of the external dosing device connected. The setting is not impor-
tant for internal dosing drives.
2.1 Connection to the MSB interface
2.1.1 Connection to Titrandos with internal dosing drive
Connect the 822 Titration Curve Simulator with the connector cable
4 to the MSB 2 connection of the Titrando.
Connect the measuring input of the Titrando with the Analog Output
7 of the 822 Titration Curve Simulator using the 6.2116.020 cable.
Adjust the communication protocol 1 to 25 bit.
Signals for the internal dosing drive can only be read at MSB 2. There-
fore you have to activate the checkbox Send dosing signals to
MSB 2 under System/Diagnosis/822 Curve Simulator. This
setting will only be deactivated when restarting the PC Control software
or switching off and on the Touch Control.