5 Appendix

36

Spectrosense, Instructions for Use

Carry out adjustment

Carry out the voltage adjustment as described in Section 5.2.2.

Titrando with tiamo

Connect the Spectrosense to your Titrando (see Section 2.1) and start the
tiamo software.

Enter Spectrosense into sensor list

If the Spectrosense has not yet been inserted into the sensor list you
can do this now in the program part Configuration under Sensors.

Click Edit, New and choose Other sensor.... Enter a Sensor name
and choose the Device and the Measuring input. Confirm the entry
with [OK].

Create method

Switch to the program part Method. With File, New you can create a
Blank method and confirm your selection with [OK].

With Insert, New command enter a new measuring command
MEAS U and confirm with [OK]. Doubleclick the measuring com-
mand in order to open the corresponding properties window.

Adapt the following parameters in the tab General/Hardware:
Device

name

connected Titrando

Measuring input

1 or 2

Sensor

previously entered Spectrosense

Stirrer off

In the tab Measuring parameters change these parameters:
Measurement without drift control activate

Measuring time 999'999

s

Measuring interval 0.5

s

In order to be able to start the method you first have to save it with
File, Save. Choose the Method group in which you want to store the
method, enter a Method name and click [Save].

Start measurement

Switch to the program part Workplace. In the sub window Run under
Sample data select the method you previously created. The meas-
urement can now be started with [Start].