4 Maintenance

30

Spectrosense, Instructions for Use

4 Maintenance

4.1 Storage

• Rinse the Spectrosense with dist. water between the titrations.
• Rinse the Spectrosense thoroughly with dist. water and dry it before

storing it.

• Use only soft tissues to wipe the mirror surface and make sure no for-

eign material is on it, as this would involve the danger of the surface
getting scratched.

4.2 Troubleshooting

Problem

Remedy

There is no
output signal.

• Connect the Spectrosense to the MSB connection

of your Titrando or to the stirrer connection of your
Titrino.

• Connect the Spectrosense to the measuring input

'Ind.' of your Titrando or Titrino.

• Remove the Spectrosense from the solution and

check whether there is light at the tip. If you cannot
see any light you have to replace the Spec-
trosense. The LED cannot be exchanged.

The output signal
is noisy or
unstable.

• Check if the mirror is tightly screwed onto the

Spectrosense.

• Check if there are any foreign materials or air

bubbles in the optical path.

• Check whether the connection to the power supply

exists. The connection must not be under me-
chanical stress.