3.11 Inserting the heating tube into the KF titration cell

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

874 Oven Sample Processor

6.1830.030

E.3010.032

6.2730.020

1

2

3

Figure 15

Volumetric KF titration cell

Proceed as follows:

1 Guide the upper part of the septum stopper over the tip of the

6.1830.030 heating tube as shown in the illustration.

2 Guide the E.3010.032 O-ring over the heating tube.

This O-ring is also part of the 6.1244.040 set of seals that is enclosed
with each KF titrator. It cannot be reordered individually.

3 Guide the lower part of the septum stopper over the heating tube

and screw the three parts together (not too tightly).

4 Insert the stopper with the tip of the heating tube into the front

opening of the KF titration cell

5 Shift the height of the tip of the heating tube vertically in such a way

that the outlet opening of the tubing is immersed as deeply as possi-
ble. The tip must not however be permitted to get in the way of the
stirring bar in the KF titration cell. Afterwards give the septum stop-
per its final tightening.